Glassware
How to choose the appropriate sample vessel for your sample
|
Sample Aggregation?
Liquid
|
|
|
|
Solutions, suspensions, blood, plasma, cell suspension, etc
Suitable sample vessels:
Capillary
- For aqueous solutions maximal volume 50 µl
- Used for liquid samples in solvents with high dielectric constant like e. g. water, alcohol, DMF
- Standardized and calibrated sample vessels made of borosilicate
- Cheap one way items
- Easy sample exchange if used together with a matching guidance tube
 

Sample Tube
- In dependence of the dielectric constant and the polarity of your solvent you can use sample tubes with an outer diameters of 3 to 6 mm, length 135, e. g. 4 to 6 mm sample tubes for radicals in hydrocarbons, benzole, toluole
|
 |
 

Flat Cell
- Used for samples in solvents with a high dielectric constant. The ideal sample distribution in the resonator leads to a 2 - 2.5 fold higher intensity compared to 50 µl capillaries.
- You can choose between flat cells with capillary tubes or normal tubes and sealing with ground neck collets or closing caps.
- The crucial point, when using a flat cell is the orientation of the flat section, the vertical position and the filled volume of the flat cell. Therefore we recommend using the special holder SH-P as the system to achieve always the same sample positioning.
 

|
| |
Solid
|
|
Powder, solid compounds, vessels, tissue
Suitable sample vessels:
Sample Tube
- The humidity of the sample restricts the outer diameter of the sample tube you can use.
- Usually 4 and 5 mm sample tubes are used
- In the case of high sample humidity you can use a finger dewar for measurement (see below).
- The length 240 mm is adapted to the finger dewars, we offer.
 

Fixed temperature dewar
- In the case of high sample humidity like in vessels and tissue you can either freeze the sample put it into a sample tube and measure it in a fixed temperature dewar at 77 K or you can use the frozen cylinder directly for the measurement using the dewar.
 
|
|
Special cuvettes
- This cuvette is used alanine dosimetry of thin film test strips. The cuvette has a flat section too.
- Furthermore we offer special tubes for alanine tablet dosimeters.
 

Emulsion, paste, skin sections, tissue sections
Tissue cell
- The tissue cell is especially developed for skin sections, tissue sections, paste and emulsions.
- You can clean it very easily.
- The crucial point using a tissue cell is the orientation of the flat section, the vertical position and the filled volume of the flat cell. Therefore we recommend using the special holder SH-P as the system to achieve always the same sample positioning.
 

For detailed information about the glassware we offer, please  click here and download our catalogue.
|
 |
|
| |
|
|

 |
 |
 |
 |
 |
 |