Benchtop EPR Spectrometer MiniScope MS100

Detectable Species for Quantification
  • Free radicals
  • Paramagnetic transition metal ions
  • Metals
  • Crystal lattice (defects)
Samples
  • Aggregation from liquid to solid,
    e. g. liquids, solutions, suspensions, blood, cells, paste, vessels, tissue, powder, solid compound
Temperature of the Sample
  • Resonator temperature (32 - 34°C)
  • - 196 °C (Accessories: Dewar, liquid nitrogen)
  • - 170 to + 200 °C for capillaries and sample tubes of outer diameter 3 mm (Accessory: Temperature Controller TC H02)
Detection Limit
  • 25 nM TEMPOL in 50 mM phosphate buffer, pH 7,4 *
  • 2 x 10 Spins/0,1 mT

* Sample vessel capillary, microwave power 20 mW, modulation amplitude 0,2 mT, scan time 4 times 60 s, time constant 100 ms

Packages for Special Applications
  • Alanine dosimetry
  • NO-Fe(II)DETC
  • Reactive oxygen species (ROS)
Packages for Automation
  • Flow through system, Multi-Channel
  • Flow through system, One-Channel

Sample Vessels
  • Capillaries, sample tubes (outer diameter 3 - 6 mm)
  • Flat cells (Width of the flat section 5 or 7 mm)
  • Tissue cell
  • Dewar
  • Special cuvettes

1 µM TEMPOL, 50 mM phosphate buffer pH 7.4*

Accessories
  • Temperature Controller TC-H02
  • Auto sampler
  • General accessories
  • Glassware
  • Software
Delivery
  • Spectrometer
  • Guidance Tube F 120/1.5 for capillaries
  • 250 pieces 50 µl capillaries
  • Sealing kit for capillaries
  • Standard-Software
  • MiniScope Control
    (control software):
    • Adjust of the measurement parameters
    • Automatic or manual tuning
    • Recording of Spectra
    • Saving of data
    • 4096 data points per spectrum, etc.
 
  • Analysis (calculation software):
    • Amplitude calculation
    • Integration
    • Baseline correction
    • Filtering
    • g-value estimation
    • Summation of spectra, etc.
  • Multiplot (presentation software):
    • Overlay of spectra
    • Display of spectra in series
    • Adaptation of the gain
    • X-Axis title selection: data points, field or time
    • Export of data in ASCII files, etc.

Extensions for the MiniScope MS100 you can find in the Catalogue (EPR Spectrometer)

 

Technical Data

   
Computer System
Processor   from Pentium II upwards
Operating system   Microsoft Windows 98 SE, 2000 und XP
Interface   RS232 or USB

 

Microwave Bridge
Microwave power   100 µW - 50mW
Resonator   rectangular TE102
Microwave frequency   9.3 - 9.55 GHz
Tuning   automatic
 
Electro Magnet
Modulation amplitude   5 µT - 0.7 mT
Modulation frequency   100 kHz
Magnetic field range   50 - 450 mT (optional - 50-600 mT,
water cooling required)
Field sweep   0 - 400 mT (optional 0 - 550mT, water cooling required)
Sweep time   12 s - 34 min
Stability   2 µT
Homogeneity   ± 5 µT in the sample region
Cooling   air (optional water)

 

Case
Size   53 x 42 x 36 cm
(length x width x height)
Weight   about 60 kg
Power consumption   about 500 W
 

magnettech GmbH
Analysenmesstechnik
Louis-Blériot-Str. 5, D-12487 Berlin
phone:  +49 30 6780 2526
fax: +49 30 6322 4101
e-Mail:  info@magnettech.de
magnettech


 
© 2007 Magnettech