RadicalScope mt 500-L (L-Band, 1 - 1.5 GHz)

Detectable Species for Quantification
  • Free radicals in vivo or in samples with high water content
Spatial Signal Resolution
  • Spectral, spectral-spatial, 2D imaging and 3D imaging in dependence of gradient systems for spatial signal resolution
Object of Investigation
  • Mouse, rat, skin, tooth, tissue
Temperature of the Sample
  • Room temperature
Resonators
  • Surface coil resonator, with automatic matching control, compensating movements e. g. breathing of an animal, coil diameter 8 and 10 mm

  • Re-entrant resonator, highest sensitivity, suitable for measurement in the tail or head of a rat, accessible diameter 18 mm
Detection Limit
  • Depends on type of resonator and the filling factor
  • Re-entrant resonator: 1 µM signal to noise ratio 5 : 1, TEMPOL in 50 mM phosphate buffer, pH 7,4

Spatial Resolution of EPR Signals
  • Spatial signal resolution 0.1 mm for spectral spatial and 1 mm for 2D and 3D imaging in dependence on the line width of the signal

Three 200 µl capillaries orientated and filled the way shown above.The y-axis gives the spatial information about the distance; the x-axis shows the spectral information.

 
Delivery
  • Magnet especially designed for L-band
  • Magnet power supply
  • Microwave bridge
  • Power supply for control and microwave bridge
  • Computer and 17'' TFT monitor
  • Standard-Software
    RadicalScope Control
    (control software):
    • Adjust of the measurement parameters
    • Manual tuning
    • Recording of Spectra
    • Saving of data
    • 4096 data points per spectrum, etc.

Extensions for the RadicalScope mt 500-L you can find in the Catalogue (EPR Spectrometer)

 

Technical Data

 
  • Analysis (calculation software):
    • Amplitude calculation
    • Integration
    • Baseline correction
    • Filtering
    • g-value estimation
    • Summation of spectra, etc.
  • Multiplot (presentation software):
    • Overlay of spectra
    • Display of spectra in series
    • Adaptation of the gain
    • X-Axis title selection: data points, field or time
    • Export of data in ASCII files, etc.
Computer System
Processor   from Pentium II upwards
Operating system   Microsoft Windows 98 SE, 2000 und XP
Interface   PC controlled by Sorcus board

 

Microwave Bridge
Microwave power   100 µW - 100 mW
Microwave frequency   1 - 1.5 GHz
Control   Control Unit CU5
Tuning   Manual tuning

 

Magnet Power Supply
Stability   10-6/min
Output power   500 W

 

Optional Resonators

Surface coil resonator
Re-entrant resonator

 
Electro Magnet
Modulation amplitude   0.5 µT - 0.5 mT
Modulation frequency   100 kHz
Magnetic field range   0 - 65 mT
Field sweep   0 - 50 mT
Sweep time   10 s - 2 h
Stability   2.5 µT/min
Homogeneity   10-5/2 cm3
Free gap size
(without modulation and tomography coils)
  150 mm
Weight   500 kg
Cooling   air, water in the case of 1D to 3D extensions

 

Extensions for Imaging
Gradient   3 mT/cm
 

magnettech GmbH
Analysenmesstechnik
Louis-Blériot-Str. 5, D-12487 Berlin
phone:  +49 30 6780 2526
fax: +49 30 6322 4101
e-Mail:  info@magnettech.de
magnettech


 
© 2007 Magnettech